Replacing boards / replacing components
6 - 9
Siemens AG
Register 5
RXR2-130.061.02
Page 9 of 18
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
Checks and adjustments
6
•
Check the I.I. electrode voltages according to the test protocol for the I.I.
- Do not readjust in cases where there are only slight deviations from the values on the
test protocol (measurement device tolerances).
•
Check that the camera optics are centered to the I.I. output and adjust them if necessary;
refer to "Replacing the camera optics ", subitem "Checking the centering of the camera
optics to the I.I. output " and "Centering the camera optics to the I.I. output".
•
Check the camera rotation and readjust it if necessary.
•
Check the setting of the radiation iris and readjust it if necessary.
•
Check the dose rate and readjust it if necessary.
•
Perform the IQ Quick test.
- When doing this, check the resolution first and, if necessary, readjust the optical
resolution of the Compact optics.
•
Complete the country-specific acceptance. (§16 Partial acceptance... /DHHS...)
Replacing the compact optics
6
Removing the Compact optics
6
•
Turn the C-arm so that the I.I. is at the top.
•
Loosen and remove the I.I. cover.
•
Rotate the camera in the "0" position (6/Fig. 6).
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.
Fig. 5
Fig. 6
WARNING
3
6
3
2
5
4
5
1
5
1
2
4
4
3
5
3