25 AI 2P
60 AI 2N
26 AI 3P
61 AI 3N
27 AI 4P
62 AI 4N
28 NC
63 NC
29 NC
64 GND
FM-07
30 CANL
FM-09
65 CANH
FM-08
31 NC
66 NC
32 GNDPWR NC
67 NC
33 GNDPWR FM-06
68 P24V
FM-03
34 GNDPWR FM-05
69 P24V
FM-02
35 GNDPWR FM-04
70 P24V
FM-01
The following applies to the connections:
NC
Not connected
x/3
Normal closed
x/2
Center
x/1
Normal open
FM-
XX
XX indicates the number present on the core of the connecting cable.
Notes on the inputs/outputs:
● Digital outputs (DO): with changeover contacts, freely parameterizable, e.g. for measuring
range identification
Load: 24 V AC/DC/1.7 A, isolated, non-sparking
● Digital inputs (DI): designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
● Analog outputs (AO): 0/4 … 20 mA, isolated; max. load 750 Ω
● Analog inputs (AI): 0/4 … 20 mA, non-floating, internal resistance ≤ 100 Ω
Note
Display of connections AO1..3 on the LUI
The connections identified by AO1..3 are displayed on the LUI as AO11..13.
Note
Strain relief of the connecting cables
Make sure that the strain relief of the cable gland is not overloaded. Make sure that the
connecting cables are force-free.
If necessary, install an external strain relief.
Procedure for connecting
1. Open the center cover of the field control unit.
2. Open the cable gland using suitable tools.
Wiring
6.5 Field control unit
Field device
Operating Instructions, 05/2015, A5E35640463-02
61
Summary of Contents for SIPROCESS GA700
Page 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Page 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Page 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Page 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Page 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...