All inlets/outlets are provided with flame arresters.
Note
Condensation in the analyzer module
Use a sample gas cooler that reduces the dew point of the sample gas to the extent that the
gas does not condense in the analyzer. The internal sample gas paths are not heated. The
sample chamber is temperature-controlled at 72 °C.
Note
Formation of condensation in the module enclosure
The module enclosure is vented by the breathing device. Any condensation formed in the
enclosure is discharged by the breathing device.
Note
Installation/removal of the analyzer module
It is not envisaged that you install the analyzer module in the field module enclosure or remove
it from the enclosure. If necessary, replace the complete field module. Contact the responsible
servicing technician.
Description
3.2 Field module with OXYMAT 7 analyzer module
Field device
26
Operating Instructions, 05/2015, A5E35640463-02
Summary of Contents for SIPROCESS GA700
Page 6: ...Inhaltsverzeichnis Feldgerät 6 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 64: ...Anschließen 6 5 Feldbediengerät Feldgerät 64 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 66: ...Bedienen 7 1 Bedienen Feldgerät 66 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 112: ...Index Feldgerät 112 Betriebsanleitung 05 2015 A5E35640463 02 ...
Page 118: ...Table of contents Field device 6 Operating Instructions 05 2015 A5E35640463 02 ...
Page 176: ...Wiring 6 5 Field control unit Field device 64 Operating Instructions 05 2015 A5E35640463 02 ...
Page 178: ...Operation 7 1 Operation Field device 66 Operating Instructions 05 2015 A5E35640463 02 ...
Page 222: ...Appendix A 6 Returned delivery Field device 110 Operating Instructions 05 2015 A5E35640463 02 ...
Page 224: ...Index Field device 112 Operating Instructions 05 2015 A5E35640463 02 ...