Adjusting the Max. Brightness/Contrast
SIMOMED HM
RA52-060.061.01.07.02
Siemens AG
07.02
TD SD 21
Page 22 of 24
Medical Solutions
Table for H/C Values
NOTE
Adjust the values that have been set.
The gradations, 5% in the 0% field, as well as 95% in the 100% field
must be visible; if necessary, use the complete tolerance range!
Tab. 3
B/C Value
System
Max.
Back-
ground
(cd/m
2
)
max. Contrast
Brightness / Contrast
Adjustment using the
front buttons or in the
SSW
Max.
Contrast
( cd/m
2
)
Max.
Brightness
( cd/m
2
)
Contrast (
cd/m
2
)
Brightness
( cd/m
2
)
Important: see
1
Card Systems (prior
to AXIOM)
ACOM
1. The monitor with Part No. 74 11 726 is not used in this application (max. luminous intensity
= 450 cd/m2)
3 ± 0,3
560 ± 30
1 ± 0,2
500 ± 20
0,8 ± 0,1
Important: see
1
Multispot TOP
Neurostar TOP
Angiostar TOP
Multistar KK
3 ± 0,3
560 ± 30
1 ± 0,2
400 ± 20
0,5 ± 0,15
Polystar TOP
Fluorospot H Sys-
tems
Fluorospot TOP
(1k/2k)
Fluorospot Compact
AXIOM Artis MP
3 ± 0,3
400 ± 20
1 ± 0,1
260 ± 10
0,5 ± 0,15
Magic VIEW
=> in the 90% or
20% field of the
SMPTE test pattern!
3 ± 0,3
229 ± 5
9 ± 0,5
229 ± 5
9 ± 0,5
Leonardo (WS)
3 ± 0,3
560
± 30
1 ± 0,1
260 ± 10
0,5 ± 0,15
AXIOM Artis
FA / FC / BA / BC
3 ± 0,3
450 ± 20
1 ± 0,1
400 ± 20
0,5 ± 0,1