Configuration with Monitor Replacement
17
Page 17 of 24
Medical Solutions
Installation of the cover panels is in the reverse of the above instructions.
The terminal resistors for the inputs on
the video amplifier can be changed using
jumpers.
•
X4 => BAS input terminal resistor.
- 1- 2 => ON
- 2-3 => OFF
- Standard (setting when shipped)
ON
•
X5 => H- Synch input terminal resis-
tor.
- 1- 2 => ON
- 2- 3 => OFF
- Standard (setting when shipped)
OFF
•
S1 cable compensator.
- 1 => ON
- 2 => OFF
- Standard (setting when shipped)
OFF
Fig. 6:
Video terminal resistor / cable
compensator