10
Service User Interface
SIMOMED HM
RA52-060.061.01.07.02
Siemens AG
07.02
TD SD 21
Page 10 of 24
Medical Solutions
4-
3Service User Interface
Initial Installation of the Service Program
0
•
Insert the supplied service diskette in the disk drive.
•
Select drive A in the Windows file manager.
- Double-click on the "setup.exe" file to start the program; the installation routine will in-
stall the service software in the service window. If a service window is not yet in-
stalled, it will be installed automatically.
- Select the language (English or German), select the Continue button.
- Confirm the installation from drive A by clicking on Continue.
- Confirm installation after C:SIM_HM by clicking on Continue.
- Confirm the Service program group by clicking on Continue.
- Close the window which displays successful installation by clicking on OK.
Starting Installation of the Service Software
0
•
Installation of the Service Software is started in the Windows service window by dou-
ble-clicking on the monitor symbol.
•
Com 1 is selected as the default port. If a connection is not made, the Com port can be
reselected in a window which appears, or the off-line mode can be selected.
NOTE
For an explanation of the masks or of individual functions in a
mask, a help text can be selected in the masks by clicking on a
question mark button.