Medical Solutions
Troubleshooting / Repair / Replacement
0
Beginning / since February 2001, the monitor is only replaced completely when there is a
malfunction.
•
Every defective monitor must be returned with an exact description of the malfunction!
•
Without a description of the malfunction, sporadic and/or temperature-related effects or
even system-dependent causes almost never can be found.
NOTE
If a middle monitor in an MTS 1 / 3 / 6 needs to be replaced, the po-
sition of the monitor to the rear longitudinal strut can be changed
so that the monitor can be replaced without “spending extra time”
on it.
However, the document
for MTS 1/3/6
(AX52-X01.812.01 / Transverse Strut Horizontal Rota-
tion at the Bottom)
for MTS 6, upper traverse
(AX52-X01.812.01 / Turning the Top Hor-
izontal Transverse Strut)
must be observed!
WARNING
Increased risk of injury!
If the installation document is not observed, the traverse can fall
and crushing can occur.
¹
Installation Instructions, AX52-X01.812.01.... absolutely
must be observed.
Test Equipment and Aids
0
•
Laptop with the Windows operating system.
•
Serial interface cable (0 modem cable).
See CB - DOC TD00-000.801.01.... (Spare Parts Catalogue),
Part No. 99 00 440 RE999.
NOTE
A gender changer is no longer needed.
•
Monitor Multinorm Testbox.
See CB - DOC TD00-000.801.01.... (Spare Parts Catalogue)
Part No. 97 16 754 Y4905
•
Set of Torx offset screwdrivers.
See CB - DOC TD00-000.801.01.... (Spare Parts Catalogue)
Part No. 99 00 663 GE999