Siemens AG
RA52-060.061.01.07.02
SIMOMED HM
07.02
TD SD 21
Configuration with Monitor Replacement
13
Page 13 of 24
Medical Solutions
•
Ambient light sensor on means: if the ambient light sensor is activated, the contrast of
the monitor is readjusted according to the ambient light. Control is in the range between
the contrast setting (front control panel) and max. contrast setting in the SSW.
NOTE
The behavior of the ambient light sensor can be adjusted.
Adjustment is made in the SSW under: => Adjustment => Ambient
light sensor
We always use the “Default 1” values in this screen.
Norm Change Time
0
In this window, the delay time when there is a norm change can be adjusted.
Function: if the current video norm is changed, the monitor waits the time that has been
set before it starts a norm search routine and again correctly displays the new norm. If
there are instable video sources, “unwanted” norm changes and image flicker can be pre-
vented by extending the norm change time.
In the SSW, select under: => Configuration => Scan Update time
Size / position and 1:1 / 3:4 format relative to the application.
0
Size / Position
NOTE
The size / position must be adjusted under => Adjustment => Ge-
ometry
correctly displayed.
It must be ensured that a square is also displayed square.
Example, SMPTE test pattern:
The 5 horizontal boxes must have the same display length as the
5 vertical boxes on the screen (measure using a ruler, etc.).
1:1 / 3:4 Format
In this window, the format musts be manually configured to 1:1 or 3:4 for certain applica-
tions
(see the list - blocked Norm
). Manual selection is required because the monitor
cannot detect the format from the video signal.
In the SSW, select under: => Configuration=> Disable Scan
Application
Power Down
Ambient Light Sensor
All (except Magic View)
off
see corresponding IQAP / IQ specifications
Magic View
on
off