TD SD 21
Adjusting the Max. Brightness/Contrast
19
Page 19 of 24
Medical Solutions
Adjustment
0
NOTE
Adjustment sequence:
The basic setting must have been performed!
The brightness / contrast adjustment is performed using the front
buttons or in the SSW.
The bypass adjustment must be performed only if the “emergency
mode” and thus generally for a different norm for the “normal
mode”
NOTE
Ambient light sensor:
The ambient light sensor is switched off automatically by the SW
when the window is opened for the adjustment of the brightness /
contrast basic setting and for bypass.
When "Brightness / Contrast Adjustment using the front buttons
or in the SSW" is performed, the ambient light sensor must be
switched off in => Configuration => Power down & Ambient light
sensor.