Application Planning
3.2 Application and ambient conditions
Fail-safe operation of the Mobile Panel 277F IWLAN
Function Manual, 08/2008, 6AV6691-1FQ01-2AB0
43
Testing for mechanical environmental conditions
The following table provides information on the type and scope of tests to determine
mechanical ambient conditions for the HMI device.
Tested for
Test standard
Comments
Vibrations
IEC 60068, part 2–6
(sinusoidal)
Type of vibration:
20 frequency cycles with a tuning rate of
1 octave/minute.
Frequency range:
10 ≤ f ≤ 150 Hz, ± 1 Hz
Deflection:
0.35 mm / 5 g ± 15% at the control point
Shock
IEC 60068, part 2–27
Shock form: Half-sinus
Shock amplitude: 30 g
Shock duration: 11 ms
Number of shocks: 3 per axis
Continuous shocks IEC 60068, part 2–29
Shock form: Half-sine
Shock amplitude: 10 g
Shock duration: 16 ms
Shock cycle: (1–3)/s
Number of shocks: 1000 ± 10
Impact
IEC 60068, part 2–75
One-time impact stress of 1 Nm with an impact
test device similar to DIN VDE 0740, Part 1,
Section 19.2 at room temperature.
Falling
Drop testing in accordance
with
EN 60068-2-32
1.2 m
Applies to the HMI device with and without
battery:
Reducing vibrations
If the HMI device is subjected to greater shocks or vibrations, you must take appropriate
measures to reduce acceleration or amplitudes.
We recommend fitting the charging station of the HMI device to vibration-absorbent material
(on metal shock absorbers, for example).
Climatic ambient conditions for the HMI device
The following table shows the permitted climatic ambient conditions for use of the HMI
device:
Ambient conditions
Permitted range
Comments
Temperature
•
Operation
•
Storage/transport
•
0 to 40 °C
•
–20 to 60 °C
Relative humidity
5 to 85 %,
no condensation
Corresponds to relative humidity, load
degree 2 in accordance with IEC 61131,
part 2
Atmospheric pressure
1060 to 700 hPa
Corresponds to an elevation of –1,000 to
2,000 m