Clearing Blockages
Clearing a Blockage in the Measurement Block Tube
11. Disconnect the measurement block tube from the sample pump tubing
connector.
Figure 3-47. Disconnecting the Measurement Block Tube
12. Thread the clot removal line up the measurement block tube until it
appears in the measurement block. Pull the line through.
Figure 3-48. Clearing a Blockage in the Measurement Block Tube
13. Carefully inject water into the sample pump tubing, until water
appears at the waste cap.
Figure 3-49. Clearing a Blockage in the Sample Pump Tubing
14. Reconnect the sample pump tubing to the measurement block tube.
Maintaining the System
Page 3-33