Page 2-8
248 Operator’s Manual
Measuring a Short Sample or a Sample with a Bubble
The 248 detects short samples or bubbles in samples and gives you 2
options:
•
reposition the sample so there are no air bubbles under the sensors and
then continue with the analysis, or
•
flush the sample out of the measurement block ready to repeat the
analysis.
The 248 beeps and displays:
OR
This message is displayed for 1 minute. If no action is taken the 248 flushes
the sample out of the measurement block.
Continuing with the Sample Analysis
1.
If you want to measure the sample lift the front cover and look at the
measurement block.
2.
Turn the sample pump rotor (the sample pump is the lefthand pump)
in the direction indicated, so that the sample is repositioned directly
beneath the sensors for which results are required.
Figure 2-2. Repositioning Samples
CAUTION:
Do not move samples backwards as KCl from the reference
sensor may contaminate the other sensors.
Short sample
press
*
to cancel
Reposition sample then
press
#
to measure
press
*
to cancel
Reposition sample then
press
#
to measure
Bubble in sample