Page 3-24
248 Operator’s Manual
Replacing the Probe and Tubing and Probe Housing
BIOHAZARD:
See Appendix A,
Protecting Yourself from Biohazards
.
Equipment: Probe and tubing kit (Cat. 673251) or Probe and housing
kit (Cat. 673253), as required; disinfectant (for example, Cat. 673390).
1.
Use the
Disinfect
routine and stop the 248 system, Page 3-8.
2.
Raise the front cover.
3.
Lift the probe lever to the second position.
4.
Push the probe connector to the left and pull it out of the reagent
inlet connector.
Figure 3-28. Disconnecting the Probe Connector
5.
Lift the probe lever past the second position and hold in place.
Figure 3-29. Probe Lever Positions
Closed
First Position
Second Position
Holding past the second position
Probe Connector