Page 2-2
248 Operator’s Manual
Analyzing Syringe Samples
BIOHAZARD:
See Appendix A,
Protecting Yourself From Biohazards
.
1.
If the main message shows capillary or QC sample press
1
to select
syringe.
2.
Lift the probe lever to the second position.
3.
Slide the syringe sample onto the probe and gently push the probe
sleeve back. The 248 beeps when the probe sleeve is in the correct
position and starts sampling.
CAUTION:
Try to position the probe to obtain the most representative
sample – do not allow the probe tip to touch the syringe plunger.
NOTE:
If you have a very small syringe sample, present the sample
and press
*
, then confirm syringe sample when prompted.
4.
Hold the sample in place. The 248 beeps when sampling is complete.
5.
Remove the sample and close the probe.
Lift probe to analyze
syringe sample
11:03:48
press
#
for menu
Ready
or press
1 capillary
2 QC
Probe open
Present sample
Sampling
close probe to cancel
Sampling complete
Moving sample
Please wait
press
*
to cancel
close probe to cancel
Hold sample
in place
Remove sample
and close probe