Maintaining the System
Page 3-9
Stopping the 248 System
1.
From
Ready
press
#
for
menu
,
2
for
Maintenance
and
5
for
Stop
System
.
2.
Carry out the maintenance task.
3.
Press
*
to restart the 248 system.
The 248 calibrates on return to the
Ready
screen.
Using the Prime Routine
This routine drains and pumps solution through the 248 and should be
used when replacing the pump tubing, changing reagents or pumping
disinfectant through the manifold.
Draining the 248
1.
Raise the front cover and remove the Buffer pack and Wash bottle.
Do not remove the Waste bottle. Place tissues under the bottle
tubes to catch any spillage.
2.
From
Ready
press
#
for
menu
,
2
for
Maintenance
and
3
for
Prime
.
3.
The 248 pumps and drains the system.
Priming the 248
1.
Follow the appropriate maintenance procedure, then from
Ready
press
#
for
menu
,
2
for
Maintenance
and
3
for
Prime
.
2.
The 248 primes the system.
If necessary, repeat the routine to thoroughly prime the system.
3.
Press
*
twice to return to the
Ready
screen.
Press
*
to restart
→
Maintenance
→
Prime
Priming
please wait
→
Maintenance
→
Prime
Priming
please wait
System stopped