Precision and Recovery on Whole Blood
Whole blood was tonometered at 37
°
C and run on four 248 pH/Blood Gas
Analyzers.
Analyte n
WRSD
Expected
Observed
%Recovery
%CV
Level 1
pH
15
0.008
7.502
7.491
99.9
0.10
p
CO
2
15
0.435
21.3
21.7
102.0
2.00
p
O
2
15
2.819
376.3
368.5
97.9
0.77
Level 2
pH
17
0.006
7.377
7.371
99.9
0.09
p
CO
2
17
0.454
35.4
35.6
100.5
1.28
p
O
2
17
1.139
85.1
84.4
99.2
1.35
Level 3
pH
20
0.008
7.300
7.299
100.0
0.11
p
CO
2
20
0.522
49.6
49.0
98.7
1.07
p
O
2
20
0.301
49.8
49.8
100.0
0.60
Level 4
pH
19
0.010
7.073
7.064
99.9
0.14
p
CO
2
19
2.081
99.1
98.9
99.8
2.10
p
O
2
19
2.309
148.6
147.4
99.2
1.57
Level 5
pH
19
0.007
6.972
6.972
100.0
0.10
p
CO
2
19
0.737
148.6
148.5
99.9
0.50
p
O
2
19
1.252
106.2
104.8
98.7
1.19
WRSD = within-run standard deviation
Precision on Controls
Data was collected from nine runs across four 248 pH/Blood Gas Analyzers
over nine days.
pH
Level
n
Mean
WRSD
TotSD
%CV
1
108
7.141
0.002
0.006
0.08
2
108
7.403
0.002
0.009
0.12
3
108
7.610
0.002
0.012
0.15
Page E-2