Installing the Gas Cylinders
WARNING
Compressed gas cylinders require careful handling. To prevent
damage and possible personal injury, observe the following precautions:
Compressed gas cylinders require careful handling. To prevent damage and
possible personal injury, observe the following precautions:
• Never drop cylinders, allow them to strike each other or subject
them to other strong shocks.
• Secure cylinders to a wall or bench, or place in a cylinder base
support stand.
• Never drag, roll or slide cylinders, even for short distances. Use a
suitable hand truck to move cylinders.
• Never tamper with safety devices in regulators or cylinders.
• Use these gases for the calibration of clinical and research
instrumentation only. US Law prohibits dispensing these gases for
drug use.
• The contents are under pressure – do not puncture.
• Do not use or store near heat or open flame.
• Do not expose cylinders to temperatures above 54
°
C (130
°
F) as this
may cause the contents to vent or explode.
• Never throw cylinders into fire or incinerators. Follow the disposal
instructions on the cylinders.
1.
Place the gas cylinders into their final position and secure.
2.
Remove the plastic protective wrapping from the valve assemblies.
3.
Fit the gas cylinder seals on the regulators.
4.
Attach the gas regulators to the cylinders with the regulator nipple
engaging the opening in the cylinder valve.
5.
Attach and tighten the yoke screws firmly.
6.
Attach and tighten the tubing adapter fitting into the needle valves.
7.
Connect one end of the lengths of cylinder connecting tubing to the
regulator fittings.
8.
Slowly open the cylinder valves until the pressure gauge indicator
stops rising (approximately 3/4 of a turn, average psi is 2000 -
minimum working pressure is 300 psi). Open the cylinder valves one
more turn.
CAUTION:
The needle valve is easily damaged. Turn the valve gently
when opening.
9.
Open the needle valve fully and allow the gas to flow for a few seconds.
10. Connect the tubing from the calibration gas cylinder to the cal gas
connector tubing on the rear of the 248. Connect the tubing from the
slope gas cylinder to the slope gas connector tubing on the rear of the
248. Make sure there are no kinks or other restrictions in the lines.
11. Listen carefully for any gas leakage. Use soapy water to check for
bubbles.
12. Check that the regulator outlet gauges read 4
±
1 psi.
13. Lower the front cover.
Installation
Page F-5