Control LEDs
MAMMOMAT Balance
SPB7-115.840.01 Page 7 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
Control LEDs
CPU 01-170
_________________________________________________________________________________
Color
Function
LED1
YELLOW
Don’t care
LED2
YELLOW
blinking during x-ray
LED3
YELLOW
bright during x-ray ( mAs meter pulses)
LED D1
RED
IRQ activity monitor, blinks in normal condition; if bright fix CPU
hanged UP
LED DU1
YELLOW
in2,5Vdc
LED DU2
YELLOW
in3,3Vdc
LED DU6
RED
ON at start UP during FPGA initialization
LED DU8
RED
ON during CPU initial reset
LED DU11
YELLOW
+5Vdc
CPU Reset sequence
At power ON, LED DU6 and LED DU8 are bright, LED DU8 goes OFF, LED DU6 goes OFF, LED
D1becomes bright fix and after blinks to indicate that CPU is running.