Troubleshooting
MAMMOMAT Balance
SPB7-115.840.01 Page 18 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
ALARM
TUBE THERMAL LIMIT REACHED
Code
HU
When
Starting exposure
Why
X-ray tube assembly has reached maximum operating temperature.
Action
Wait for tube cooling.
Evaluate operating conditions from last 1300 exposure memory.
Check tube temperature sensor PCB 02-171.
Disable tube rotation brake.
ALARM
TECHNICAL ERROR
Code
Refer to list
When
Any time.
Why
Internal error having no meaning for the operator.
Action
Check Log File to find out what it is.
Date and time of error / fault can be helpful to locate it quickly
2) Technical menu
When
Any time.
Why
Broken Fuse or defective block has been located by CPU diagnostic functions
Action
Check Log File to find out when it occurred
SIEMENS AG
CPU 01-170 F2=OK F4=OK
SW1 00000000 SW2 0000
INVERTER Vdc (in) = 545 V LV= OK
Filament 92-073 >> PASS <<
Rotating Anode 92-074 >> PASS <<
FIRMWARE REV BAL__ XX TUBE
Tube Housing kJ
25°C