DIP switches
MAMMOMAT Balance
SPB7-115.840.01 Page 6 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
DIP switches
PCB 01-170 Main CPU
SW1
DIP1 =
ON
Anode current calibration mAH
DIP2 =
ON
Anode current calibration mAL
DIP3 =
ON
DEMO mode without x-ray
DIP4 =
ON
Pulsed radiation disabile with grid table
DIP5 =
ON
AEC frozen to single Film/Screen
DIP6 =
OFF
Not in use, must remain OFF
DIP7 =
ON
Compression force calibration
DIP8 =
ON
Mo/Rh filter enabled
SW2
DIP1 =
ON
C-arm Brake, disabled for service
DIP2 =
ON
Inverter Vdc input sense disabled
DIP3 =
ON
Max compression force 200N
DIP4 =
ON
Exposure start on left to right grid movement
DIP5 =
OFF
Future expansion
PCB 92-083 Keyboard
Language
DIP1
DIP2
DIP3
DIP4
Italian
OFF
OFF
OFF
OFF
French
ON
OFF
OFF
OFF
English
OFF
ON
OFF
OFF
German
ON
ON
OFF
OFF
Spanish
OFF
OFF
ON
OFF
Polish
ON
OFF
ON
OFF
Turkish
OFF
ON
ON
OFF
Portuguese
ON
ON
ON
OFF