Filament
MAMMOMAT Balance
SPB7-115.840.01 Page 29 of 62
Siemens AG
Troubleshooting Guide
Rev.02 10.04
Medical Solutions
To enter
mAL
current calibration set
DIP2 of
SW1
( PCB
01-170) = ON
.
In the mAs window of the display, the following parameters will appear:
•
mAH
or
mAL
, filament current
I
f
in Ampere;
•
mAs value selected before entering calibration mode
•
after exposure , expected anode current value and exposure
I
a
in
mA
.
After exposure the following condition can be reached:
a) Anode current less than 70% of the expected value
in place of the anode current value, on the display
LLL.L
will appear.
It means that exposure ended by means of Back-Up Timer.
Increase
I
f
filament
current by 0.050 and repeat the exposure.
If the display indication is still
LLL.L
, repeat the above point till you get the indication of mA value
and no more the
[LLL.L]
b) Anode current higher than 10% above expected value
in place of the anode current value, on the display
HHH.H
will appear.
Decrease
I
f
filament current by 0.050 and repeat exposure.
If the display indication is still
HHH.H
, repeat the above point till you get the indication of mA and
no more the
HHH.H
.
mAH
I
f
4,710
A
90
I
a
90,1
mA
mAs
40