ADJUSTMENT
6.4.5
One-point adjustment using the keypad (function CAL1)
When adjustment conditions (gas pressure, humidity and CO
2
concentration) are different
from the operating conditions of the TRANSIC151LP, set the ambient parameters of the
TRANSIC151LP to the adjustment environment for the adjustment duration. These settings
must then be changed back to represent the process conditions when the TRANSIC151LP
is reinstalled in its operating environment.
see “Setting ambient parameters”, page 40
further information on setting ambient parameters for the TRANSIC151LP.
1
Check that no error messages are active.
Active error messages influence adjustment. Malfunction messages,
2
Enter the password in menu
PAS
.
3
Select menu item
Cal1
. This freezes the analog output.
4
Connect the reference gas.
5
Enter the known O
2
value and confirm with
Ent
.
6
The measured value display blinks.
7
Feed calibration gas.
8
Wait until the display shows a stable value.
9
Confirm with
Ent
.
PASS
is displayed when the calibration is successful. The TRANSIC151LP now calculates
the new gain and offset settings and starts to display the new measured value.
10
Press
Ent
twice.
This ends the one-point adjustment.
Fig. 38: One-point adjustment using the keypad.
6.4.6
Restoring factory calibration
Restoring the TRANSIC151LP factory calibration via the keypad,
tory calibration (FAC)”, page 38
Back
Back
k
<
<
<
<
<
<
<
!
Ent
Ent
2 x
<
<
Ent
Ent
Dn
Dn
Up
Up
Ent
Ent
Ent
Ent
<
5 s
The adjustment can be aborted at any time with
Back
.