INSTALLATION
3.1.3
Strong light sources near the oxygen measuring probe
The interfering effect is influenced by:
●
Filter used
●
Light incidence angle into the measuring probe
▸
Use a stainless steel mesh filter for ambient light, e.g., normal interior or laboratory light.
▸
Use a PTFE filter for especially strong light sources, e.g., outdoors with direct sunlight.
3.1.4
Pressure
Observe the specifications on pressure conditions,
see “Flange mounted TRANSIC151LP-
“TRANSIC151LP-C with sample gas cell and wall bracket”, page 13
and
“Pressure suitability”, page 78.
NOTE:
Strong light sources interfere with TRANSIC151LP operation
▸
Prevent strong light sources reaching the measuring probe.