O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
OPERATION
4
4.5.8
Scaling the analog output (ASCL)
The analog output can be scaled freely.
3
After pressing
Ent
, use submenu item
Set Lo
to set the oxygen value to be transferred
with the lower mA value (4 mA or 0 mA).
4
Use submenu item
Set Hi
to set the oxygen value to be transferred with the upper mA
value (20 mA).
Fig. 27: Scaling the analog output
4.5.9
Digital output (ALA)
1
Press
Ent
to display the current switching position.
2
To check the switching function, press
Ent
and use
Dn
and
Up
to select the desired
switching function
OPE
(open) or
CLO
(closed).
Fig. 28: Checking and changing the digital output state
4.5.10
Resetting the measuring device (rESE)
TRANSIC151LP is restarted.
Fig. 29: Restart of the TRANSIC151LP oxygen sensor
4.6
Maintenance interface
Observe the Operating Instructions delivered with the serial interface cable.