O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
4
OPERATION
4.4.3
Calibration gas, actual value (CAL.C)
1
Freezes the analog output.
2
Displays the current O
2
concentration measured.
Fig. 18: Calibration gas, actual value display
4.4.4
Signal intensity(SIL)
1
Compares the current signal intensity of the laser on the receiver against the signal
intensity from the factory calibration.
2
The signal intensity serves to determine contamination on the optics. Important: The
laser signal can be amplified so that values above 100% are possible.
Fig. 19: Signal intensity display
Information on alarm thresholds,
see “TRANSIC151LP behavior when errors occur”,
.
4.4.5
Display of active and undeleted errors (ERR)
This menu displays all active error messages.
describes how the errors are read out
and deleted. The display shows
E 0
when all errors have been deleted. Significance of error
numbers,
Fig. 20: Displaying all errors currently existing