8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
ADJUSTMENT
6
6.2.2
Using bottled gas
●
Preparation for calibration with bottled gas
see “Setting up the gas supply for calibra-
Gas flow adjustment
.
●
When calibration conditions (gas pressure, humidity and CO
2
concentration) are
different from the operating conditions of the TRANSIC151LP, set the ambient
parameters of the TRANSIC151LP to the adjustment environment for the adjustment
duration. These settings must then be changed back to represent the process
conditions when the TRANSIC151LP is reinstalled in its operating environment.
●
Let the gas flow in.
●
Wait until the measured value has stabilized.
●
Now compare the TRANSIC151LP display value with the calibration gas specification.
●
Set the parameters for pressure, humidity and temperature corresponding to the
process conditions.
●
Ensure the analog output is no longer frozen.