Operating Instructions
ISD 230/260/280
Chapter
4
© SICK AG · Auto Ident · Germany · All rights reserved
8 008 207/30-11-2001
19
Installation
4.3
Mounting and adjusting the system
4.3.1Mounting the ISD on the three-point bracket
Risk of damage to the housing!
The maximum depth of engagement of the two nuts M 5 is
10 mm
from housing surface.
Longer screws will damage the housing.
Ø
Use screws with the correct length.
1.
Prepare the two three-point brackets for either vertical (angled flange left or right) or
horizontal (angled flange top or bottom) mounting.
2.
Ensure that the movable plate is parallel to the fixed plate on each bracket before they
are mounted (if necessary correct the position with the lock nuts using for the fine
adjustment).
3.
Align both three-point bracket to each other straight to the eye. The system must be
mounted in such a way that the straight line of vision between the two units is
unimpaired.
4.
Attach the two brackets at the required height with the four screws (M 6 x 12 mm) and
washers supplied using the angled flange or the fixed plate. Use the reference
markings as reference points with respect to the common reference (floor). Ensure
that the two brackets are aligned horizontally (devices must be in the same axis).
5.
Carefully insert a unit into each bracket. The two top screws on the rear of the housing
facing the cover (optical adjustment aid) must fit into the locking holes of the movable
plate so that the bottom of the device (connector) is flush with the bracket (
Fig. 4-3
).
Check: the reference marking (horizontal) must be at the same height as the two glass
lenses.
Fig.
4-3: Design of the three-point bracket
Locking hole
Lock nut
w/f 10
Reference marking (vertical)
Angled
flange
Reference marking (horizontal)
Movable
plate
Fixed
plate
Unit mounted on
three-point bracket
ATTENTION