68
FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH
Product Description
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
2.2.6
Meter body option (spoolpiece)
The FLOWSIC100 Flare can also be fitted with an optional spoolpiece to reduce geometric
uncertainty of device installation and to simplify assembly. The exact design (nominal
diameter, connection, material) always depends on the customer specifications.
The installation length of the spoolpiece depends on the nominal diameter of the pipe:
– Installation length 800 mm for pipe diameter up to 28”
– Installation length 1100 mm for pipe diameter 30” ... 60”
– Installation length for pipe diameter >60‘‘...72‘‘ on request
System solutions (FLOWSIC100 Flare + spoolpiece) are available as options with pressure
and temperature transmitters. For the positioning of pressure and temperature
transmitters the following configurations are available:
– Spoolpiece, standard length with integrated pressure tapping, temperature sensor 3D
downstream (customer pipeline)
– Spoolpiece, extended length with integrated pressure and temperature tapping.
Fig. 28
Spoolpiece option (example)
800 mm