4.2.2
Entering application specific parameters for single-path configuration in the
MCUP
Select device file "MCU-P" and move it to the "Project " window.
Set the MCUP to "Maintenance" mode and enter the Level 1 password (
).
Select the "Parameter / Measuring Point 1(2/3) / Measuring Point Parameter" direc-
tory.
Enter the data in the "Measuring Point 1 - Application Parameter (I)" group as listed in
the following table.
Fig. 102
"Parameter / Measuring Point 1 / Measuring Point Parameter" directory
Entry field
Parameter
Remark
TAG
Name
TAG No. of the measuring point
Description
Name
Description of measuring point
Pipe diameter
Value
Pipe diameter in m
Nominal diameter
(inch)
Value
Select the corresponding value
Sensor type
Installation version Select the corresponding type (cross-duct or probe)
Adjust factor
Value
Has to be determined by a comparision measurement;
without measurement enter "1"
Low flow cut-off
Value
Minimum velocity of gas to be suppressed
Suppress negative
speed
Active
A negative gas velocity is suppressed
Damping time (T90)
Value in s
Response time of measured value at selected measuring
point
NOTICE:
For flow calibrated devices the coefficients CC0 … CC4 must be the same as
those detailed on the calibration certificate. If necessary, set the parameters in
SOPAS ET with values from the calibration certificate.