166
FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH
Start-up and Parameter Settings
Su
bj
ect
to
ch
an
ge
with
ou
t n
otic
e
Select the "Configuration / Application Parameter" directory (
) and
enter the values determined for path length and angle determined in Section 3.3.1.4 as
well as the cross-sectional area.
Fig. 101
Configuration / Application Parameters" directory (example for settings)
Installation angle
Angle between measuring axis and main gas flow direction (path
angle
α, →
Path length
Distance transducer - transducer (path length L,
Cross-sectional area
Internal diameter of the pipeline (
)
●
The parameters entered are saved in the FLOWSIC100 Flare after the
switching from "Maintenance" to "Measurement".
●
Set installation parameters are converted automatically when the unit
of measurement is changed.
●
Enter the calibration coefficient for gas velocity Cv_1 ("Calibration Coef-
ficients" group) as a negative value if FLOWSIC100-EXPR is installed at a
vertical pipeline (
).