8.3 Outputting the Process Variable
8.3.1 Specifying the Process Variable Output
TS
0
(CR)(LF)
Once this command is received, the recorder transfers the process variable to the transmission
buffer when the (ESC) T command is received.
8.3.2 Updating the Data
(ESC)T(CR)(LF)
Once this command is received, the recorder transfers the updated process variable to the
transmission buffer.
8.3.3 Specifying the Process Variable Output Order (For BINARY Mode Output)
BO
0
(CR)(LF)
Outputs from the upper byte.
BO
1
(CR)(LF)
Outputs from the lower byte.
The recorder outputs from the lower byte, unless otherwise specified.
8.3.4 Outputting the Data
FM
0
, S CH, E CH (CR) (LF)
Specifies the ASCII mode output.
FM
1
, S CH, E CH (CR) (LF)
Specifies the BINARY mode output.
S CH: Output start channel
01 to 02
E CH: Output end channel
01 to 02
When the number of channels is one, specify (S_CH) = (E_CH).
(e.g.) FM0, 01, 02(CR) (LF) ASCII, Ch1 to Ch2 output