112
11.1.3 Structure
Mounting: Panel mounting (vertical panel)
Allowable backward inclination: Within 30
Material (Color)
Case: Polycarbonate (Black), Glass 10 % UL94-V0
Door: Polycarbonate UL94-V2 (Transparent)
Dust-proof, drip-proof (IEC60529-IP65)
11.1.4 Power Source
Rated supply voltage range:
100 to 240 V AC
Working supply voltage range: 85 to 264 V AC
Rated power frequency:
50/60 Hz
Working frequency range:
45 to 65 Hz
Table 11.1.4 Power consumption
100 V AC
Max. power consumption
1 pen
Approx. 20 VA
Approx. 30 VA
2 pens
Approx. 25 VA
Approx. 35 VA
11.1.5 Normal Operating Conditions
Ambient temperature: 0 to 50
Ambient humidity:
20 to 80 %RH
Supply voltage:
85 to 264 V AC
Supply frequency:
45 to 65 Hz
Vibration:
10 to 60 Hz 0.2 m/s
2
max.
Shock:
Unacceptable
Magnetic field:
400 A/m max. (DC and AC: 50/60 Hz)
Noise:
Normal mode (50/60 Hz)
DC voltage, Thermocouple: Peak value lower than the span width
of the measurement range
RTD: 50 mV max.
Common mode (50/60 Hz): 250 V AC max.
Interchannel maximum noise voltage: 200 V AC at 50/60 Hz
Posture:
Allowable backward inclination: Within 30 , Right-and-left level
Warm-up time:
30 minutes minimum from the time of power-on