103
9.3 Replacing Consumables
Replace consumables according to the following list.
No.
Name
Type
Period
Remarks
Quantity
1
Chart paper
H-10100
33 days
When chart speed is 20
mm/h
1
2
Ribbon cassette
HPSR001H0002C
6 months
Logging printing every 3
hours
1
3
Cartridge pen
WPSR196A000001A 2 months
1 pen, about 2km in a beeline
1
4
Cartridge pen
WPSR196A000002A 2 months
2-pen, about 2km in a beeline
1
9.4 Adjusting the Pen Recording Position (Point Calibration)
The following describes how to adjust pen position on the chart paper.
It is advised to adjust recording position annually for accurate recording.
How to Adjust
Leave the recorder as it is, and adjust it according to the instructions in Section 7.2.11 Point
calibration. (See p.90)
9.5 Calibration (Data Calibration)
Calibrate a process variable input every year in order to keep measurement accurate.
Devices required for calibration
Precision voltage (current) generator (Jointly use a digital voltmeter with accuracy of
0.02 %
or less as required)
Precision dial resistor
[Note]
(1) Over time ribbon colors will begin to fade. To ensure clear recording, replace the ribbon
cassette as soon as possible.
(2) Use the ribbon cassette within one year after its purchase. Over time, ribbon colors will begin
to fade.
(3) If color is noticeably faint before the changing period, replace the cartridge pen.
[Note]
Prior to calibration, allow warm-up for at least 30 minutes after switching the power ON.