VM-7 System Manual
Chapter 2 Description of the System
15
No. Name
Function
⑥
2ch monitor output
connector
Outputs 2ch input signal.
There are the following three methods to output
signals, and the signal to be output is set through
FIELD CONFIG.
y
Outputs input signals through the buffer amplifier
y
Outputs after the input signal (speed vibration) is
integrated and converted into the displacement
vibration.
y
Outputs after the signal is calculated using 1ch
signal and converted into the absolute vibration
(displacement).
⑦
3ch monitor output
connector
Outputs 3ch input signal via the buffer amplifier.
⑧
4ch monitor output
connector
Outputs 4ch input signal.
There are the following three methods to output
signals, and the signal to be output is set through
FIELD CONFIG.
y
Outputs input signals through the buffer amplifier
y
Outputs after the input signal (speed vibration) is
integrated and converted into the displacement
vibration.
y
Outputs after the signal is calculated using 3ch
signal and converted into the absolute vibration
(displacement).
⑨
Module fastener
This is used to fix the Absolute Vibration Monitor
Module to the instrument rack.
⑩
Main board
A board on which electronic parts are mounted
⑪
DIN connector
A connector used for connecting to the motherboard