8 Maintenance
8-14
it into the waste tank.
(4) Connect the new drain tube to the leveler as described in "8.2.3 Removing the Leveler".
(5) Insert the drain tube into the tube hook. Ensure that there is no slack in the drain tube between
the leveler and the tube hook.
(6) Secure the drain tube and the leveler exhaust tube with the tube-holding block. Ensure that there
is no slack in the drain tube between the tube hook and the tube-holding block.
(7) Insert the drain tube into the waste tank. Ensure that the end of the drain tube is not immersed in
waste.
(8) Fill the leveler with water.