6 Specifications
6-3
6.5.4 User Support
x
Development
assistant
x
Diagnosis
assistant
6.5.5 Instrument Control
x
Monitor display of instrument status
x
Control of vacuum pump
x
Automatic igniting/extinguishing of plasma
x
Automatic extinguish function that operates when analysis completed
x
Automatic control of autosampler
6.6 Utilities
6.6.1 Power Supply Requirements
Single phase,200/220/230/240 V ±10% (upper limit: 250 V), 50/60 Hz, 30 A
6.6.2 Ground
30
Ω
max., independent ground
6.6.3 Argon Gas
Purity
99.95% min.; Pressure: 450±10 kPa
6.6.4 Cooling water
Mains water
Pressure: 100 kPa to 300 kPa; Flow rate: 1 L/min min.
CCD cooling water Temperature setting range: 0
°
C to room temperature
Temperature control precision: ±0.1
°
C
Cooling capacity: 200 W min.; Liquid temperature: 10
°
C
Pumping capacity (maximum pump head): 2.1 m min. at 50 Hz, 2.4 m min. at
60 Hz
Coolant may be pumped to the main unit only while pumping signals are
being output from the main unit.
Pumping signals: 24 VDC, 0.5 A
6.6.5 Installation Conditions
Temperature
18
°
C to 28
°
C (temperature must not fluctuate more than 2
°
C/hour)
Humidity
20% to 70% (with no condensation)
6.6.6 Dimensions and Weight
Dimensions
1,380 mm (width) × 740 mm (depth) × 760 mm (height)
(not including plasma stand exhaust guide)
Weight
270
kg
6.6.7 Options
Autosampler
Radial-view observation kit (option)
Hydrofluoric acid sample introduction system
Ultrasonic nebulizer
Summary of Contents for ICPE-9000
Page 2: ......
Page 6: ...iv...
Page 14: ...xii 1 3 2 Warning Labels...
Page 16: ...xiv...
Page 20: ...xviii...
Page 26: ...1 Introduction 1 2...
Page 64: ...5 Configuration and Operating Principle of the ICPE 9000 5 26...
Page 74: ...7 Operation 7 6...
Page 106: ...8 Maintenance 8 32...
Page 114: ...9 Maintenance Parts and Consumable Items 9 8...