5 Configuration and Operating Principle of the ICPE-9000
5-14
5.14 Radial-View Observation Kit (Option)
This optional unit is installed inside the plasma stand and enables radial-view observation. Switching
between axial-view observation and radial-view observation is possible from the software. The
observation points at the top and bottom set in radial-view observation can also be switched from the
software. At the time of installation, the points 9 and 15 mm (for the optional plasma torch) above the RF
coil are set but these can be changed with the observation position setting knobs. (Setting range: 7 to 22
mm.)
Radial-view
observation kit
Observation direction indicator
Upper observation
position setting knob
(15 mm)
Bonnet
Radial-view
observation purge tube
Observation height
indicator
Lower observation position setting knob
(9 mm)
Locking knobs
Fig. 5-11 Radial-View Observation Kit
WARNING
Risk of Burn Injury
There is a risk of burn injury. When checking the plasma stand, do NOT touch the
plasma stand interior for 3 minutes after the plasma is extinguished.