41
10. Calibration
10.2.2 Span Calibration With External Weights
“E-CAL” (The balance will be adjusted with external calibration weights)
1
From the mass display, press the
[CAL]
key once.
If the display shows “E CAL” (Preset calibration
method is “E-CAL”), jump to Step
4
.
If the display shows another item (Preset calibra-
tion method is not “E-CAL”), go on to Step
2
.
for preset calibration method)
2
Press the
[CAL]
key repeatedly until “FUnC.SEL”
appears, then press the
[O/T]
key to display “CAL”.
3
Press the
[O/T]
key. The display shows “E CAL”.
4
While “E-CAL” is shown, press the
[O/T]
key.
Span calibration begins and the zero display blinks.
5
Recheck that there is nothing on the pan and
wait.
6
The set weight value (Refer to
) appears,
blinking.
7
Place the weight indicated on the pan. Wait
until the blinking zero display reappears. (It may
take about 30 seconds.)
8
Remove the calibration weight. After “CAL End”
appears for several seconds, the display returns
to the mass display and the span calibration is
complete.