5. Basic Operation (Read Chapters 1 to 5 for basic but proper operation of the balance.)
For Basic Operation
5.2 Changing the Unit Display
Pressing the
[UNIT]
key switches display between the registered units, piece counting and specific
gravity measurement modes. The units other than 'g', 'pcs', '%' and 'ct' are not registered in the default
settings.
5.3 Switching the Weighing Range
(AUW-D series only)
In the dual-range balance AUW-D series, when the power is connected and turned on, the balance is
set to the “small range” with a minimum display of 0.01mg. To switch to the “large range” with a mini-
mum display of 0.1mg, press the
[1d/10d]
key (except for AUW-D series, this key has a different func-
tion.
) . When measurement exceeds the small range capacity (82g for AUW220D, 42g for
AUW120D) during use of the small range, display automatically switches to the large range. In this
case, taring with
[O/T]
key in the large range will fix the range, and reducing the load on the pan within
the small range capacity will not return it to the small range.
[O/T]
key has to be pressed again, within
the small range, to resume the small range display.
Note
The units to be used must be registered as described in 12.Units. If the
power is turned off and turned on again, the weighing unit will be 'g'. (The
registered units remain.)
Small range
Large range