40
10. Calibration
10.2 Executing Calibration
Perform calibration only after correct installation and thorough warm up. Also, make sure that nothing
is on the pan and ensure conditions free from the influence of vibrations or wind.
10.2.1 Span Calibration With Built-in Weight
(AUW-D/AUW/AUX series only)
“i-CAL” (The balance will be adjusted with the built-in calibration weight.)
1
From the mass display, press the
[CAL]
key once.
If the display shows “i-CAL” (Preset calibration
method is “i-CAL”), jump to Step
5
.
If the display shows another item (Preset calibra-
tion method is not “i-CAL”), go on to Step
2
.
for preset calibration method)
2
Press the
[CAL]
key repeatedly until “FUnC.SEL”
appears, then, press the
[O/T]
key to display
“CAL”.
3
Press the
[O/T]
key again. The display shows
“E CAL”.
4
Press the
[CAL]
key twice. The display shows
“i-CAL”.
5
Press the
[O/T]
key. The motor sound of weight
loading system is heard and the display changes
automatically in this order: “CAL 2”, “CAL 1”, “CAL
0”, and “CAL End”.
When the display returns to the mass display,
span calibration is complete.
...