67
13. Application Functions
10
Place the items to be measured on the
immersed hanging pan. The display shows the
weight value in water. Press the
[CAL]
key to
show the specific gravity value in the set specific
gravity value display mode.
Press the
[POWER]
key to return to 9. If air bub-
bles adhered to the item being measured and the
result is a specific gravity value that differs from
the expected value, you can recalculate the spe-
cific gravity value by retrying the weight measure-
ment in water after removing the air bubbles from
the item.
11
To make the next measurement, unload the
pans, press
[CAL]
, and begin again at Step 5.
When finished measuring specific gravity, press
the
[UNIT]
key.
(Example)
(Example)
(Hold display mode example)
Notes
• Although the density display shows 4 decimal places, not all of these
digits will stabilize depending on the conditions.
• When the items to be measured are placed in the liquid, they must be
completely immersed and must not have bubbles adhering to the sur-
face of the sample or the immersed hanging pan.