10. Calibration
10.2.3 Calibration Check With Built-in Weight
(AUW-D/AUW/AUX series only)
“i-tESt” (The balance is checked with the built-in weight but not adjusted.)
1
From the mass display, press the
[CAL]
key once.
If the display shows “i tESt” (Preset calibration
method is “i-tESt”), jump to Step
5
.
If the display shows another item (Preset calibra-
tion method is not “i-tESt”), go on to Step
2
.
for preset calibration method)
2
Press the
[CAL]
key repeatedly until “FUnC.SEL”
appears, then, press the
[O/T]
key to display
“CAL”.
3
Press the
[O/T]
key again. The display shows
“E CAL”.
4
Press the
[CAL]
key repeatedly until the dis-
play shows “i tESt”.
5
Press the
[O/T]
key. The display changes auto-
matically in this order: “tESt 2”, “tESt 1”, and “tESt
0”. After that, the display shows the drift from the
previous span calibration for several seconds.
6
After “tEStEnd” appears for several seconds,
the display returns to the mass display, and the
calibration test is complete.
Note that the Calibration Test procedure DOES
NOT adjust the balance, it only reports the drift
from the previous span calibration.
...
(Example)