10. Calibration
10.3.5 Inputting External Calibration Weight Value for E-CAL
The exact value (conventional mass value) of the operator’s calibration weight to be used for E-CAL and
E-tESt procedures can be entered.
1
From mass display, press the
[CAL]
key repeat-
edly until “SEttinG” appears. Press the
[O/T]
key.
“CAL dEF” appears.
2
Press the
[CAL]
key repeatedly until “CAL SEt”
appears and press the
[O/T]
key. One of these
numbers will appear, depending on the model:
50.0000g, 100.0000g, 200.0000g, or 300.0000g.
In the upper part of the display panel, the
symbol and the
symbol appear in order to
indicate numerical input mode. The leftmost
numeral blinks. The blinking digit may be
changed.
3
When the
[UNIT]
key is pressed, the value of
the blinking digit increases by 1 at a time. Press
the
[PRINT]
key to move the blinking digit one
place to the right. Enter the desired value.
4
Press the
[O/T]
key to set the external calibra-
tion weight value. The display proceeds to “SEt”
and then to “CAL SEt”.
5
Press the
[POWER]
key again to return to the
mass display.
Calibration Weight Value Range
(Example)
(Example)
AUW-D series
220D
75g to capacity
120D
35g to capacity
Other series
320
145g to capacity
220
95g to capacity
120
45g to capacity
Note
In the AUW-D series the conventional mass value may be input for up to
five decimal places.