61
SI 580E DIN SIZE WEIGHING CONTROLER
www.sewhacnm.co.kr
7-2.
Calibration Process
Display
Cause
Treatment
Eerr-01
When Max.capacity/digit value is
over 20,000
Re-input the Max. Capacity, less than 20.00
(Max. Capacity / Digit)
Eerr-04
Standard weight value is over than
Max. Capacity
Re-input Standard weight value with Number
keys, under Max. Capacity
Eerr-05
Standard weight value is less than
10% of Max. Capacity
Re-input Standard weight value with Number
keys, more than 10% of Max. Capacity
Eerr-06
1. Amp. Gain is too big
2. Sig+ and Sig- wire connection
error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and
real weight, please re-input the value
(set value is too small)
Eerr-07
1. Amp. Gain is too small
2. Sig+ and Sig- wire connection
error
3. Test weight is not loaded
Check standard weight’s weight with set value.
If there is difference between set value and
real weight, please re-input the value
(set value is too big)
Eerr-08
Under “F-function” model, set value
is “N.A”
Check the correct value and re-input
err-a
When there is continuous vibration
on the weighing part,, indicator
cannot process calibration any
more.
- Find vibration cause and remove
- Load cell check
- Load cell cable and connecting condition
check