22
SI 580E DIN SIZE WEIGHING CONTROLER
www.sewhacnm.co.kr
F-FUNCTION LIST
General Function Setting (“●” Factory default set value)
Weighing Data Save Method selection
F01
●
0
Non-Save Mode (Weight Data & Frequency)
1
Save Mode ( Weight Data & Frequency)
Weight –Back up selection
F02
0
Normal Mode
●
1
Weight Back up Mode
Motion Band Range setting
F03
05 01~99
This is set “Steady” acceptable range of weighing part.
If there is vibration on weighing part, you can set this function and reduce
the vibration effect on weighing process.
1 : Weak vibration ~~ 99 : Strong Vibration
Zero Tracking Compensation Range setting
F04
05 00~99
Due to external causes (Temperature, wind, and dust), there will be small
weight difference, the Indicator will ignore the weight difference and display
as Zero.
Auto Zero Range setting
F05
00 00~99
Within the “Auto Zero” range, weighing part is steady, indicator will display
current weight as “Zero”
If the weighing part is not “Steady”, indicator will display current weight.
(Auto Zero Range : ± Set value + weight unit)
Digital Filter setting
F06
04 00~40
Weak vibration (0) ~ Strong Vibration (40)
Zero key Operation mode selection
F07
0
Activate under “Steady” condition, only
●
1
Always activate
Zero key Operation Range selection : (-) value is same to (+)
F08
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
●
2
Activated within 10% of Max. Capacity
3
Activated within 20% of Max. Capacity
4
Activated within 50% of Max. Capacity
5
Activated within 100% of Max. Capacity
6
No limit of Zero key operation range.
CAUTION : If setting over than 10%, The display weight could be over than Load cell input
signal or Max. Capacity and it may display “CELL-Err” or incorrect weight value. And It can