SI 580E DIN SIZE WEIGHING CONTROLER
www.sewhacnm.co.kr
7-3.
Digital Weighing Indicator
Display
Cause
Treatment
“
Cell
-
er
”
or
“
0Uer
”
1. Load cell Error
2. Load cell cable Error
3.Load cell connection Error
4. A/D Board Error
5.If Analogue value
is over 1,040,000.
※ When weigh “-“ value,
If it is over set max capa, “OVER” is
displayed.
Ex) Even though set max capa is “100”
and it is over “-100”, “OVER” is
displayed.
1. Under “TEST” mode 1, check
analogue value. If you cannot
get any analogue value or there
is no change although adding
load, please check load cell,
load cell cable, connection
conditions first.
2. Replace another load cell, and
check the indicator condition. If
you have same problem, please
replace new indicator and check
A/D board error.
3. Try to connect the indicator’s
A/D with the other indicator.
4. Check the power and connection
of terminal.
“
UNpa55
”
1. Power is ON, when some materials
are on weighing part.
※ Under “Normal Mode”, if there are
more than 20% loading of Max.
capacity, “Un-Pass” display will be
appeared and indicator will stay
until removing the load.
※Setting Back-up mode it can
memory empty value, and it
becomes set value without
displaying” Un-pass”)
1. If you set “Normal Mode”,
please check weighing part
empty or not before turn on the
power. If there are some
materials in/on weighing part,
please remove those materials
and turn on the power.
2. Please try to set F02-01(Back-up)
mode so that the indicator can
remember first empty value.
“
5et
”
When Power is on, “SET” displays.
It means EEPROM has some problem.
Please contact the distributor or
Head Office.
“
Halt
”
H/W has some problem.
“
t-err
”
The dead Battery
※ Under
“
Cell
-
er
”
, Zero key, Tare key, Hold key and print key will not be activated.