23
SI 580E DIN SIZE WEIGHING CONTROLER
www.sewhacnm.co.kr
be the cause of load cell damage.
Tare key Operation Range selection: (-) value is same to (+)
F09
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
●
2
Activated within 50% of Max. Capacity
3
Activated within 100% of Max. Capacity
“Hold” Mode selection
F10
●
0
Peak Hold : Measure Max. weight value and hold on display.
1
Sample Hold : Hold current weight until “Hold Reset”
2
Average Hold : Hold average value (Refer F-F54)
“STEADY” condition check time setting
F11
03
00
∫
99
During the set time period, estimate weighing part’s “STEADY” condition and
display.
If you set small value, indicator will take “STEADY” fast, if you set value,
indicator will take “STEADY” slow. ( 0.5sec per set value)
Display Up-Date speed setting
F12
●
1
60/sec
6 6/sec
2
30/sec
7 3/sec
3
20/sec
8 2/sec
4
15/sec
9 1/sec
5
10/sec
Weight Display selection under “Unpass / OverLoad”condition
: (-) value is same to (+)
F13
0
Not Display Weight
●
1
Display Weight (with a flash)
External Input Selection 1 (IN1) – Bottom side Terminal
F14
0
Not used
7 HOLD/HOLD RESET
●
1
ZERO
8 RUN (F21-03 Under Packer Mode)
2
TARE
9 STOP (F21-03 Under Packer Mode)
3
TARE RESET
10
RUN / STOP(F21-03 Under Packer
Mode)
4
TARE / TARE RESET
11 PRINT
5
HOLD
12 PRINT the amount
6
HOLD RESET
External Input Selection 2 (IN2) – Bottom side Terminal
F15
0
Not used
7 HOLD/HOLD RESET