SERVOTOUGH SpectraScan User Manual
02400001A/1
61
10 MAINTENANCE
The Servomex SpectraScan Gas Analyzer is designed to be maintenance free for extended periods
of time. In a normal installation, it will operate unattended in its intended environment for 18 months
with no physical intervention. Typically every 18 months, the spectrometer light source needs to be
replaced.
Recommended maintenance and operation verification of the Servomex SpectraScan Gas Analyzer
consists of:
Monthly re-zeroing (performed remotely through software).
Light source assembly replacement (every 18 months or when light indicator shows < 40%
output). The replacement procedure is described in Section 10.0.
Gas cell window replacement if necessary when prompted by fault signal.