Electrical Power Requirements
The Servomex SpectraScan Gas Analyzer operates at 24 VDC or 100-240 VAC depending on the
model configuration. Be sure to adhere to the Power Rating Label on the analyzer to determine
which power source to use, and connect the correct power source.
WARNING
SYSTEM SAFETY:
This analyzer is typed as permanently connected
equipment and as such, must include a switch or circuit–breaker in the
installation. This switch must be supplied and installed by the customer (not
provided by Servomex) must be suitably located and easily reached, and must
be marked as the disconnecting device for the equipment.
WARNING
Electrical Shock and Equipment Damage:
The main circuit board in the
analyzer must have the proper fuse installed prior to connecting power.
Installing the wrong fuse in main circuit board may result in electric shock or
death and may result in damage to the analyzer. To avoid injury and
equipment damage, ensure the proper fuse is used. See Table 13 - Electrical
Specifications on page 72. Servomex provides the proper fuse and two spares
with every analyzer.
See Figure 6 - SpectraScan Interior Diagram on page 33.
Figure 7 - Example Rating Label
Summary of Contents for Servotough SpectraScan 2400
Page 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Page 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Page 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Page 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Page 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Page 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Page 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Page 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Page 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...