SERVOTOUGH SpectraScan User Manual
02400001A/1
35
6.2
Wiring
All cables entering the analyzer (24 VDC or 120/240 VAC power and communication) must be
separately jacketed as they enter the analyzer to prevent miss-wiring. The power cabling used,
whether for 24 VDC operation, or for 120-240 VAC operation, must use a 3 conductor cable. The
ground wire in this 3 conductor cable must be electrically connected to the earth ground lug in the
analyzer.
Connection
Wire type
Power (AC)
600V, 3 wire, 16AWG, Jacketed
Insulation must be rated to at least 75°C
Power (DC)
600V, 3 wire, 16AWG, Jacketed
Ethernet (TCP/IP)
CAT 5 Standard 300V
Solenoids (DC)
600V, 3 wire, 18AWG
RS485 Data
Data Cable 24 AWG, 300V
Optional Heater
600V, 3 wire, 18AWG
Table 4 - Wire Types
Summary of Contents for Servotough SpectraScan 2400
Page 7: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 v This page intentionally left blank...
Page 12: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 10...
Page 13: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 11 This page intentionally left blank...
Page 30: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 28 This page intentionally left blank...
Page 33: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 31 Figure 4 Analyzer with Cover Open...
Page 58: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 56 This page intentionally left blank...
Page 60: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 58 This page intentionally left blank...
Page 68: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 66 This page intentionally left blank...
Page 70: ...SERVOTOUGH SpectraScan User Manual 02400001A 1 68 This page intentionally left blank...